HWTS-UR007A-Herpes Simplex Virus Type 2 Nucleic Acid Detection Kit (Fluorescence PCR)
FAM: Herpes Simplex Virus Type 2 (HSV2)·
VIC(HEX): Internal Control
Step |
Cycles |
Temperature |
Time |
Collect Fluorescent Signals or Not |
1 |
1 Cycle |
50℃ |
5mins |
No |
2 |
1 Cycle |
95℃ |
10mins |
No |
3 |
40 Cycles |
95℃ |
15secs |
No |
4 |
58℃ |
31secs |
Yes |
Storage | |
Liquid |
≤-18℃ In dark |
Shelf-life | 12 months |
Specimen Type |
Female cervical swab, Male urethral swab |
Ct |
≤38 |
CV |
≤5.0% |
LoD | 50Copies/reaction |
Specificity |
There is no cross-reactivity with other STD pathogens, such as Treponema pallidum, Chlamydia trachomatis, Ureaplasma urealyticum, Mycoplasma hominis, Mycoplasma genitalium and etc. |
Applicable Instruments |
It can match the mainstream fluorescent PCR instruments on the market. Applied Biosystems 7500 Real-Time PCR Systems Applied Biosystems 7500 Fast Real-Time PCR Systems QuantStudio®5 Real-Time PCR Systems SLAN-96P Real-Time PCR Systems LightCycler®480 Real-Time PCR system LineGene 9600 Plus Real-Time PCR Detection System MA-6000 Real-Time Quantitative Thermal Cycler BioRad CFX96 Real-Time PCR System BioRad CFX Opus 96 Real-Time PCR System. |
Option 1.
Recommended extraction reagent: Macro & Micro-Test Viral DNA/RNA Kit( HWTS-3001, HWTS-3004-32, HWTS-3004-48) and Macro & Micro-Test Automatic Nucleic Acid Extractor(HWTS-3006).
Option 2.
Recommended extraction reagent: Nucleic Acid Extraction or Purification Reagent(YDP302) by Tiangen Biotech(Beijing) Co.,Ltd.