Sexually transmitted diseases (STD) remain one of the major threats to global public health security, which can lead to infertility, preterm birth, tumorigenesis, and various serious complications. There are many types of STD pathogens, including bacteria, viruses, chlamydia, mycoplasma, and spirochetes. NG, CT, UU, HSV 1, HSV 2, Mh, Mg are more common.
Reaction Buffer |
Target |
Reporter |
STD Reaction Buffer 1 |
CT |
FAM |
UU |
VIC(HEX) |
|
Mh |
ROX |
|
HSV1 |
CY5 |
|
STD Reaction Buffer 2 |
NG |
FAM |
HSV2 |
VIC(HEX) |
|
Mg |
ROX |
|
IC |
CY5 |
Step |
Temperature |
Time |
Cycles |
Fluorescence signal collection (Yes/No) |
1 |
50℃ |
5mins |
1 cycle |
No |
2 |
95℃ |
10mins |
1 cycle |
No |
3 |
95℃ |
15secs |
40 cycles |
No |
4 |
58℃ |
30secs |
Yes |
Storage:
Liquid: ≤-18℃ In dark
Shelf-life: 12 months
Specimen Type: urethral secretions, cervical secretions
Ct: ≤38
CV: ≤5.0%
LoD: 50Copies/reaction
Specificity: There is no cross-reactivity with other STD-infected pathogens such as Treponema pallidum.
Applicable Instruments: It can match the mainstream fluorescent PCR instruments on the market.
Applied Biosystems 7500 Real-Time PCR Systems
Applied Biosystems 7500 Fast Real-Time PCR Systems
QuantStudio®5 Real-Time PCR systems
SLAN® -96P Real-Time PCR Systems
LightCycler® 480 Real-Time PCR system
LineGene 9600 Plus Real-Time PCR Detection System
MA-6000 Real-Time Quantitative Thermal Cycler
BioRad CFX96 Real-Time PCR System
BioRad CFX Opus 96 Real-Time PCR System
Option 1.
Recommended extraction reagent: Macro & Micro-Test Viral DNA/RNA Kit( HWTS-3001, HWTS-3004-32, HWTS-3004-48) and Macro & Micro-Test Automatic Nucleic Acid Extractor(HWTS-3006).
Option 2.
Recommended extraction reagent: Nucleic Acid Extraction or Purification Reagent(YDP302) by Tiangen Biotech(Beijing) Co.,Ltd.