FAM: Chlamydia trachomatis (CT)
VIC(HEX): Ureaplasma urealyticum (UU)
ROX: Neisseria gonorrhoeae (NG)
CY5: Internal Control
Step |
Cycles |
Temperature |
Time |
Collect Fluorescent Signals or Not |
1 |
1 cycle |
50℃ |
5mins |
No |
2 |
1 cycle |
95℃ |
10mins |
No |
3 |
40 cycles |
95℃ |
15 secs |
No |
58℃ |
31 secs |
Yes |
Step | Cycles | Temperature | Time | Collect Fluorescent Signals or Not |
1 | 1 cycle | 50℃ | 5mins | No |
2 | 1 cycle | 95℃ | 10mins | No |
3 | 40 cycles | 95℃ | 15 secs | No |
58℃ | 30 secs | Yes |
Storage:
Liquid: ≤-18℃ In dark; Lyophilized: ≤30℃ In dark
Shelf-life: 12 months
Specimen Type: Urethral secretions, cervical secretions
Ct: ≤38
CV: ≤5.0%
LoD:
Liquid: 50 Copies/reaction; Lyophilized: 500Copies/mL
Specificity: There is no cross-reactivity for detecting other STD-infected pathogens, such as Treponema pallidum, etc.
Applicable Instruments: It can match the mainstream fluorescent PCR instruments on the market.
Applied Biosystems 7500 Real-Time PCR Systems
Applied Biosystems 7500 Fast Real-Time PCR System
QuantStudio® 5 Real-Time PCR Systems
SLAN-96P Real-Time PCR Systems
LightCycler®480 Real-Time PCR system
LineGene 9600 Plus Real-Time PCR Detection System
MA-6000 Real-Time Quantitative Thermal Cycler
BioRad CFX96 Real-Time PCR System
BioRad CFX Opus 96 Real-Time PCR System