HWTS-UR002A-Ureaplasma Urealyticum Nucleic Acid Detection Kit(Fluorescence PCR)
FAM: UU nucleic acid·
VIC(HEX): Internal Control
Step |
Temperature |
Time |
Cycles |
Collect Fluorescent Signals or Not |
1 |
50℃ |
5mins |
1 cycle |
No |
2 |
95℃ |
10mins |
1 cycle |
No |
3 |
95℃ |
15 secs |
40 cycles |
No |
4 |
58℃ |
31 secs |
Yes |
Storage | |
Liquid | ≤-18℃ In dark |
Shelf-life |
12 months |
Specimen Type | urethral secretions, cervical secretions |
Ct | ≤38 |
CV | ≤5.0% |
LoD | 50Copies/reaction |
Specificity | There is no cross-reactivity with other STD infection pathogens outside the detection range of the kit, such as Chlamydia trachomatis, Neisseria gonorrhoeae, Mycoplasma genitalium, herpes simplex virus type 1, and herpes simplex virus type |
Applicable Instruments |
It can match the mainstream fluorescent PCR instruments on the market. Applied Biosystems 7500 Real-Time PCR Systems QuantStudio® 5 Real-Time PCR Systems SLAN-96P Real-Time PCR Systems LightCycler®480 Real-Time PCR system LineGene 9600 Plus Real-Time PCR Detection System MA-6000 Real-Time Quantitative Thermal Cycler BioRad CFX96 Real-Time PCR System BioRad CFX Opus 96 Real-Time PCR System |